3-38
The following function can be performed automatically
1)
WL
Accuracy
2) WL Repeatability
3) Bandpass
4) Noise Level
5) Baseline Stability
6) Baseline Flatness
7) Hardware
8) Print Summary Report
The measure condition of GLP/GMP is shown as below:
Check item
Process
Specification
WL Accuracy
Wavelength Scan, Range: (660-650nm)
or (490.0-480.0), speed:100nm/min
+/-0.5nm
WL
Reproducibility
Wavelength Scan, Range: (660-650nm),
3 times, speed:100nm/min
+/-0.3nm
Bandpass
Wavelength Scan, Range: (660-650nm),
speed:100nm/min
3.0-4.5nm
Noise Level
Time Scan, WL:500.0nm, 5min
0.001ABS
Baseline
Stability
Time Scan, WL:500.0nm, 3600s
0.002ABS
Baseline
Flatness
Wavelength Scan, Range:1100.0-190.0,
Speed:400nm/min,
+/-0.010ABS
1. <WL
Accuracy>
After scanning the wavelength from 660.0nm to 650.0nm, the system will get the
maximum value and the corresponding wavelength position. If the wavelength
position is within the range of 656.1
±
0.5nm, the WL Accuracy item will display
<OK>.
nm
-0.001
A
GLP/GMP
1
WL Accuracy
0.0 OK
2
WL Reproducibility
3
Bandpass
4
Noise Level
5
Baseline Stability
6
Baseline Flatness
7
Hardware
8
Print Summary Report
Select Function
Sel:_
500.0