3-15
Table 3-2 Concentration Menu
Submenu
Option/Variables
Range Description
NUM WL
1 ~ 3
Numbers of Wavelength
WL Setting
190 ~1100nm
Setting Wavelength
Zoom
Rescale the spectrum of
working curve
Concentration Type 1st
k – factor
NUM STD
1 ~ 20
Number of Standard used
STD Data
0 ~ 9999
Concentration of Standards
Coff Set
K
0
K
1
C=K
1
*Abs+ K
0
Selecting Concentration Type:
2
types of Concentration calculation is available
-
1st order Linear Regression analysis
-
K- factor concentration analysis
Press 4 in the concentration type submenu, the selection of 1st order linear
regression analysis and K- factor will appear in the action bar/user prompt.
To select 1st, press 5 to define the number of standard sample for the
calibration curve. The number of standard sample that can be used in the
HALO RB-10 is 1~20.
-1st order linear
regression analysis
-K factor concentration
analysis
nm
0.001
A
1
NUM WL
1
2
WL Setting
3
Zoom
4
Concentration Type
1st
5
NUM STD
3
6
STD Data
0
End Setting
Select Function: _
500.0
Concentration
1
1st
2
k-factor