3-9
Press
Start/Stop
to begin an Autozero that use the setup parameters. The
Model HALO RB-10 will autozero automatically on every wavelength selected
in the setup before a Photometry test.
4.
Running a Photometry Test
After Autozero, the HALO RB-10 is ready for Photometry Test; data obtained
will be based on the Autozero data.
Press Start/Stop on the keyboard to begin a photometry test.
The motor will hum while moving to the required wavelength defined in the
setup. The HALO RB-10 measures the wavelength defined and displays the
following menu:
The wavelength in the left top corner will change according to the wavelength
selected when performing the test. Press Start/Stop again to measure the
next sample. The second set of data will be displayed as in the following figure.
Result of the
measurement
Display the current
wavelength position
Current data
nm
0.001
A
ID
WL
ABS
1
900.0nm
0.001
800.0nm
0.001
700.0nm
0.002
Press START Measure
1
500.0
%T/ABS