3-31
3.5.4 Post-Run Processing
Press 1 on the keyboard, and the Post-Processing commands 1-7 are
listed
The definition of the functions is outlined
1) <Zoom> Expands either or both axes
2) < Peak> Lists the peaks of a spectrum
3) <Valley> Lists the valleys of a spectrum
4) <All Data> Lists all data of wavelength scan.
5) <Smooth> Uses a smoothing routine to reduce apparent noise in
a spectrum.
6) <Derivate> Displays a 1
st
derivative spectrum.
7) <Data Reset> Goes back to the graph display based on the original
data.
5. Rescaling a Time Scan
You can expand or reduce the axis scaling of your time scan. The procedures
below that allow you to expand both the X- and Y-axis, or either one selectively.
Use the steps listed below to enter the <ZOOM> function.
Press 1 on the keyboard to come to the <ZOOM> menu.
Reenter the <Start Time> and <Stop Time> to modify the X-axis
value.
Reenter the <Upper Limit> and <Lower Limit> to modify the
500.0
nm
-0.001
A
3.000
ABS
-1.000
0
60s
Select Function:_
1
Process
2