28
Instructions for use DigiFlow
English
English
z
Turn away cover and remove battery. Ensure
proper disposal of the battery.
z
If the battery contacts are contaminated or
oxidized, clean with a dry paper towel.
z
Insert new battery. O and – symbols in
the battery compartment. The medical device is
activated automatically.
When replacing the battery, the medical device must
be checked for visible damage. If visible damage is
seen, call DrägerService.
Reassemble the medical device in reverse order.
Afterwards a function check must be performed.
Technical data
DigiFlow Standard 0 to 16 L/min
DigiFlow Neonatology 0 to 3 L/min*
02
0
021
*
at 20 °C (68 °F) and 1013 hPa
Environmental conditions
During operation
Temperature
10 to 40 °C (50 to 104 °F)
Relative humidity 10 to 90 %, without conden-
sation
Ambient pressure 800 to 1200 hPa (11.6 to
17.4 psi)
During storage
Temperature
–20 to 60 °C (–4 to 140 °F)
Relative humidity 10 to 90 % without conden-
sation
Ambient pressure 500 to 1200 hPa (7.3 to
17.4 psi)
Supply pressure
270 to 600 kPa (2.7 to 6 bar)
Flow limitation
at approx. 25 L/min (supply
pressure 500 kPa)
Display (displayable
characters)
Measured value: 7-segment
display, 2
1
/
2
characters.
Additional display: 7-seg-
ment display, 4 characters.
Bar display (12 segments, all
segments visible at
16 L/min).
Flow indicator, "low batt,"
"overflow," "timer," "min," "to-
tal," "l," "h"
Resolution of the
display
0.3 to 3.0 L/min: 0.1 L/min
higher: 0.5 L/min
Measurement
accuracy
±10 % of displayed value
Accuracy of the
display
±1 decimal place
Thread
9/16" - 18 UNF, M34x1.5 or
M32x1.5
Dimensions
B x H x T
with CS plug
56 x 144 x 140 mm
(2.2 x 5.67 x 5.51 inch)
with universal
holder
56 x 144 x 105 mm
(2.2 x 5.67 x 4.13 inch)
Weight
with CS plug
0.34 kg (0.74 lbs)
with universal
holder
0.74 kg (1.63 lbs)
Power supply
SL 760 S lithium battery, AA,
3.6 V
Battery life approx. 2 years.
Materials
Brass, aluminum, stainless
steel, plastic, gaskets latex-
free
Classification accord-
ing to Directive
93/42/EEC, Annex IX
Class ll a
Содержание DigiFlow
Страница 163: ...DigiFlow 163 164 2 low batt 174 Dr ger Dr gerService Dr gerService Dr ger Dr ger...
Страница 164: ...164 DigiFlow Dr ger EMC IEC 60601 1 2 179 O2 O2 DigiFlow...
Страница 176: ...176 DigiFlow CISPR 11 1 IEC 61000 3 2 IEC 61000 3 3...
Страница 258: ...258 DigiFlow CISPR11 1 A IEC 61000 3 2 IEC 61000 3 3...
Страница 263: ...This page intentionally left blank Instructions for Use DigiFlow 263...