14. Add-on Datalog
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
62
Mode
Start data logging by setting the mode to either stop or ring. Stop means that the
logging stops when the card is full. Ring means that when the card is full, the oldest
data will be overwritten.
Note
Never remove the SD card while the logging is running! Always set
the mode to OFF (not stop!) before removing the card. Otherwise
data loss might occur, and an error message will appear.
Note
Always use a card with sufficient storage space, even with ring
mode, otherwise the logging will not start.
Interval
The INTERVAL defines the time between subsequent data logs. You can enter an
interval of 1 second up to 24 hours.
Example:
00:00:01 means that each second the values are logged.
00:01:00 means that each minute the values are logged.
01:00:00 means that once every hour the values are logged.
Note
The instrument fills in zeros automatically.
Example: 1 is interpreted as 00:00:01 / 1: is interpreted as 00:01:00
/ 1:1: is interpreted as 01:01:00