18. Trouble-Shooting
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
78
defective or missing temperature measurement, and an exchange of the pH sensor
would be quite useless.
The raw data is also especially useful if the instrument has been hopelessly
miscalibrated.
The information on mA output values and status of the digital inputs help locating
communication problems with the control center. mA values can be measured at the
terminals of the instrument with an amperemeter.
Note
The mA outputs and the second digital input are add-ons. They are
displayed only if the functions have been activated.
18.5. Event messages
In case of trouble, the instrument displays an event message on the desktop. If there
are more than one message, they are displayed alternately.
Not all events will cause the alarm relay to switch. For many events you can decide
for yourself if the relay should switch by defining the event as an alarm action in the
alarm action list
– see Alarm relay.
18.6. Eventlog
Up to 100 events are saved in an eventlog. The *-sign indicates appearance, the
#-sign disappearance. Besides the event messages, this also lists power interrupts,
calibrations, and similar operations, with date and time.
This helps to interpret events
– for example a deviation of the measured value shortly
after a power interrupt may simply be due to the unfinished polarization of the
measurement.