12.3.2.
Touch Cal
–for instruments with touch screen
To calibrate the touch screen, select in submenu
SYSTEM=>DISPLAY the symbol TOUCH CAL to
start the calibration. The touch screen calibration
texts are available in English only.
Note
When started, the calibration routine cannot be left until it is
finished, it has to be carried out completely.
The routine comprises five steps. You have to touch each corner of the display, and
finally the center of the displayed text.
Step 1
Step 2
Step 3
Step 4
Step 5
Note
The touch screen calibration can also be started by touching the
screen for more than 30 seconds.