8. Menu Cal
– Calibration – pH
Set the controller to MAN or HOLD: MAN stops the controller, all other
functions are still active. HOLD stops the controller, all outputs are frozen to
the last value and the alarm relay is deactivated.
Mind the temperature compensation! For automatic compensation, the
temperature sensor has to be put in the calibration buffers together with the
pH sensor. If that is not possible, switch to manual mode and enter the
temperature of the buffer solutions manually.
Select the symbol CAL to enter the submenu
calibration. Here you find three symbols: run, info, and
settings.
Note
If you have selected ORP measurement, the symbol CAL is not
shown, since the ORP measurement cannot be calibrated.