8. Menu Cal – calibration
39
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
˃
With the SAVE button you store the results. With DISCARD you reject the results and keep the
previous calibration values.
Note
If the TDS factor is not between 0.1 and 5.00 saving is not possible.
8.3.
Zeropoint
˃
For Zeropoint calibration take the sensor out of the water and wipe it dry. There should be no
humidity connection between the electrodes. For plugable connection you can remove the cable
from the sensor for zeropoint calibration.
˃
In the CAL menu, select ZEROPOINT.
˃
With OK you execute the calibration and reach the info screen, showing the results of the calibration.
˃
With the SAVE button you store the results. With DISCARD you reject the results and keep the
previous calibration values.
Note
After zeropoint calibration you have to carry out a reference calibration.
8.4.
Calibration – Info
If you select the symbol INFO in the submenu CAL, you get a list of the
last calibrations. This is helpful with regard to predictive maintenance.