Use all available safety measures such as the alarm relay, the 22mA alarm current, the dosage control,
and the low-water indication.
Regularly check that all safety measures are in good working condition!
Warning
The protection built into the instrument is impaired if it is not used as intended!
1.3.
Features
Basic instrument LF
Measuring ranges
Conductivity (conductive)
Conductivity (inductive)
Temperature
0.. 2.000 µS/cm (c=0.05/cm),
0.. 20.00 µS/cm (c=0.05/cm),
0.. 200.0 µS/cm (c=0.05/cm),
0.. 2.000 mS/cm (c=0.20/cm),
0.. 20.00 mS/cm (c=1.00/cm),
0.. 200.0 mS/cm (c=10/cm)
0.. 2.000 mS/cm,
0.. 20.00 mS/cm,
0.. 200.0 mS/cm,
0.. 2000 mS/cm
-30.. +140°C
Display
Measured value and temperature with units
Additional information selectable: contact data, SD card
status, mA output, control variable, or relay status
Operation
Key pad or touch screen
Calibration
1-point calibration against reference,
zeropoint calibration possible
Measurement
Conductive conductivity (CM)
Inductive conductivity (IL)
TDS
Averaging
Off / low / medium / high, selectable via menu
Temperature measurement
Conductive: Pt100 or Pt1000, 2-wire or 3-wire connection
Inductive:
NTC 214k
Temperature compensation
Automatic or manual
(linear and for CM ulta pure water compensation)
Relay 3 – alarm relay
Potential-free contact 6A, 250 V, max. 550VA
Alarm events selectable via menu
Min. and max. limits and adjustable delay