17. Add-on Controller
87
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
17.2. S1/S2 CONFIG – configurating the controller output
Note
In this menu you define the controller output via relays. If you want to use an
analog output as controller output, you can ignore the CONFIG menu.
Instead, you have to assign the analog output to control variable S1 or S2 in
the menu IN/OUT-> mA OUT. In that case you can shut off the relay completely
in the menu CONTROL->RELAYS.
In the menu CONTROL select the symbol S1 CONFIG. Depending on your choice of controller type,
the CONFIG menu shows parameters for P /PI /PID or 3-point controller.
Note
To help you with the configuration, Neon
®
only shows those parameters
relevant for your controller choice.
Note
If you have selected an ON/OFF controller, no configuration is required, and
you cannot access the CONFIG menu.