10. Test menu
41
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
10. Test menu
The menu TEST is only accessible in the MAN Mode. In all other modes, the symbol is not shown. The
test menu always shows a list of the maximum available analog and digital outputs. Naturally, you can
operate only those outputs that are activated in your instrument. In the entry-level version this means
only relay 3.
Set the mode to manual. Select the symbol TEST in the submenu
IN/OUT.
In the submenu TEST you can switch all available relays and define mA values for the analog outputs.
To switch a relay, select ON from its drop-down list.
To check an analog output, select its settings to open the keyboard. Set the output to a value between
0.00 and 22.00 mA. This value is then sent out via the selected analog output.
Note
When you leave the menu, all relays and all analog outputs are reset to their
previous settings.