16. Add-on Modbus RTU
76
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
The alarm action list (register 6336) is coded as 32 Bit-value as follows:
Bit number
Alarm action
0
Dig. input 1
1
Dig. input 2
4
Zeropoint
5
Slope
6
Overrange
8
Lower limit
9
Upper limit
12
Dosage check
Activated add-ons
Register 2128 indicates which add-ons are activated, as recently described.
Dependent on that, some or all of the following parameters are available:
Name
1. register
Quantity Type
Bytes
Read/
write
Range
Add-on: Data log
Interval
6488
8
u_32
4
yes
yes
Mode
6496
3
u_8
1
yes
no
Logparameter Meas.
value
6504
3
u_8
1
yes
no
Logparameter
Temperature
6512
3
u_8
1
yes
no
Logparameter Raw value
6520
3
u_8
1
yes
no
Logparameter S1
7456
3
u_8
1
yes
no
Logparameter S2
7464
3
u_8
1
yes
no
Logparameter Flow
7472
3
u_8
1
yes
no
Add-on: analog outputs
mA1 22mA alarm
6368
3
u_8
1
yes
no