Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.

 

 

- 1 - 

3.7 Finishing infusion………………………………………………..…..15 

3.8 Powering off………………………..………………………………..15 

3.9 Extended operations…………………………………………….…...15 

        3.9.1 Pause/ Restart…………………………………………………15 

        3.9.2 To continue another solution container……………………….15 

        3.9.3 Purge Function..………………………………………………16 

        3.9.4 IV Set Auto-Adjustment………………………………………16 

        3.9.5 IV Set Manual Adjustment……………………………………17 

3.9.6 Occlusion valve adjustment…………………………………..18 

        3.9.7 Sound volume adjustment…………………………………….18 

        3.9.8 Toggle switch………………………………………………… 19 

        3.9.9 Battery.……..……………………………………………….....19 

3.10 Trouble Shooting……………………………………………………19 

        3.10.1 Alarms and Alarm Codes..…………………………………...19 

        3.10.2 Other failures………………………………………………...22 

3.11 Daily Check…………………………………………………………22 

Chapter IV Maintenance 

4.1 Daily Maintenance…………………………………………………...23 

        4.1.1 DigiPump LP8 enclosure……………………………………..23 

        4.1.2 Air detector……………………………………………………23 

        4.1.3 Linear pump cassette………………………………………….24 

        4.1.4 Tube clamp……………………………………………………24 

        4.1.5 Drop detector………………………………………………….24 

4.2 Battery………………………………………………………………..24 

        4.2.1 Recharging…………………………………………………….24 

        4.2.2 Replacement…………………………………………………..25 

4.3 Periodical replacement of some items………………………………..25 

4.4 Storage………………………………………………………………..25 

4.5 Explanations of symbols……………………………………………..25 

   

Warranty Terms & Conditions………………………………………………….27 

 

 

 
 
 

Содержание DigiPump LP8

Страница 1: ......

Страница 2: ... 4 1 Front view 6 1 4 2 Infusion components 6 1 4 3 LCD 7 1 4 4 Keyboard 7 1 4 5 Rear view 8 1 4 6 Bottom components 8 Chapter II Installation 2 1 Packing list 9 2 2 Installation 9 2 2 1 Set the knob bolt firmly 9 2 2 2 Set drop detector 9 Chapter III Basic Operation 3 1 Preparation 9 3 2 Installation 10 3 2 1 Install the machine 10 3 2 2 IV Set Installation 10 3 2 3 Install the drop detector 11 3...

Страница 3: ...adjustment 18 3 9 7 Sound volume adjustment 18 3 9 8 Toggle switch 19 3 9 9 Battery 19 3 10 Trouble Shooting 19 3 10 1 Alarms and Alarm Codes 19 3 10 2 Other failures 22 3 11 Daily Check 22 Chapter IV Maintenance 4 1 Daily Maintenance 23 4 1 1 DigiPump LP8 enclosure 23 4 1 2 Air detector 23 4 1 3 Linear pump cassette 24 4 1 4 Tube clamp 24 4 1 5 Drop detector 24 4 2 Battery 24 4 2 1 Recharging 24 ...

Страница 4: ...e that any applicable regulations respecting the installation and operation of the infusion pump be observed The operator should read this manual carefully and thoroughly before attempting to use the infusion pump Incorrect operation or failure of the user to maintain the infusion pump in accordance with proper maintenance procedures relieves the manufacturer or his agent from all consequent non c...

Страница 5: ...ay result Follow the instructions on this manual Set the parameters appropriately and use only the IV set which has been adjusted for the DigiPump LP8 Stop the device immediately in case of any abnormal functioning of the infusion pump Keep the DigiPump LP8 far at least 1 meter 3 0 feet from other generating HF equipments such as electrosurgical unit and others Before using the infusion pump make ...

Страница 6: ...rate 999 9ml h 10 15 20 drops ml IV set Drop rate accuracy 2 the specified accuracy is reached after the first 150 drops Flow rate accuracy 5 the specified accuracy is reached using adjusted IV set and after continued infusion over 100ml Occlusion pressure 40 140Kpa Eight levels of sensitivity available Volume infused display 0 9999ml Volume to be infused setting 1 9999ml Sound and light alarm Air...

Страница 7: ...on at operating room ICU CCU chemotherapy TPN nutrition infusion of oxytoxics and anti cancer drugs It allow the use of general purpose infusion sets Variations in drop size during the infusion flow as changed by the infusion flow is automatically corrected by a built in program in the pump to administer intravenous fluid over a long period of time without error Standard IV sets 10 15 20 60drops m...

Страница 8: ...asy single handed operation Alarm light High intensity Alarm Light Indicator can be observed at distance Handle Easy to carry Keyboard menu settings pushbuttons LCD Display with Rate and Volume Indicators infusion status and alarm messages Pump door switch Pull it to open the door Pump door Power indicator Ac and Battery Charge Indicators Linear pump cassette Air detector Occlusion sensor ...

Страница 9: ...de key Press it to start or stop the empty mode when setting infusion volume Power switch Confirm key To confirm and store the data Start Stop It lights when turn on the pump Arrow key To advance and return throughout the display menu settings Data Input Key Press it to increase or decrease the data Silence switch Press to disable and enable the audio alarm Power indicator When illuminates indicat...

Страница 10: ...nd Alarm volume AC Power receptacle Class II type CF Drip proof design AC power switch Class II type CF Drip proof design Pole Mount Auto Adjust Press the key to adjust the IV set automatically Drop detector receptacle Battery compartment Remove the screw to access the battery Battery Battery Compartment Fixation nip at the bottom when there are many infusion pumps working at the same time you can...

Страница 11: ...located on the rear of the pump 2 2 2 Set drop detector 1 As showed in the picture on the right insert drop detector plug in the receptacle at the rear panel Note If the drop detector is not well connected the display will show 2 When the drop detector is not in use it shall stay on the position showed in the picture Chapter III Basic Operation 3 1 Preparation The battery should be charged before ...

Страница 12: ...nfusion pump will perform self check automatically after it is turned on This test includes if the battery is charged if the drop detector is attached if all the system functions are normal If something is not correct an alarm will be trigged indicating the error code such as AIR LOW BATTERY Errs etc 4 After turning on the device will check the battery automatically If the charge is low it will re...

Страница 13: ...the door and pull the tube clamp D to the left to insert the IV set 5 Keep the IV set tube straight and insert it to the guides A B and C 6 Release the roller clamp Note Make sure it doesn t drip in the drop until you start the infusion 3 2 3 Install the drop detector Install the drop detector following the instructions below 1 Place the drop detector at middle point above the liquid level and bel...

Страница 14: ...t the appropriate IV set 3 4 Setting the adjustment factor Warning To keep the highest infusion accuracy we recommend to use a reliable standard PVC IV set Outer diameter is 3 4mm 4 5mm If you use one IV set which cannot meet this requirement it may affect the accuracy Possible errors that affect accuracy Error How to Solve 1 The IV set used diameter is different from the default programmed The IV...

Страница 15: ...y for reference Liquid Adjustment value Standard low surface tension 0 Dextrose 10 5 Dextrose 50 10 TPN 5 10 TPN fat soluble vitamin 10 20 1 Assume that you want to infuse dextrose 10 solution The corresponding adjustment value is 5 table 3 4 The IV set adjustment value is 3 through 3 9 4 Auto adjustment for IV set or 3 9 5 Manual adjustment for IV set 2 The total adjustment value IV set adjustmen...

Страница 16: ...e Limit Setting Range 10drops ML 1 0 999 9 1 0 166 6 1 9999ml 15drops ML 1 0 999 9 1 0 249 9 1 9999ml 20drops ML 1 0 999 9 1 0 333 3 1 9999ml 60drops ML 1 0 150 0 1 0 150 0 1 9999ml 4 Empty bottle mode When setting the Volume Limit press the The volume setting will indicate and will go to the Empty bottle mode then press to confirm Pressing start infusion and the pump will infuse all liquid in the...

Страница 17: ...peration 3 9 1 Pause Restart a Pause infusion pressing the during infusion will pause the pump b Restart pressing the again the pump will restart the infusion NOTES Pausing for more than 2 minutes will activate the alarm system Pause Overtime Press to clear the alarm and maintain the pause state 3 9 2 To continue another solution container 1 Change to a new infusion container and make sure that th...

Страница 18: ...purge function 2 Just release the to stop the purge function 3 Purge Rate for 10 15 20 drops ml IV set purge rate is 600ml h for 60 drops ml IV set purge rate is 150ml h NOTE During purge function air in line alarm will only trigger a visual alarm and not the audible alarm and the infusion will not stop The volume infused is not added to the Total Volume Infusion indicator 3 9 4 IV Set Auto adjust...

Страница 19: ...ot in the range of 30 to 30 the LCD will indicate alarm to show the failure of the adjustment process You need to turn off the pump then turn on again and restart the adjustment process 12 Check whether the solution container has been emptied If not the adjustment value will be useless Reset the machine and repeat the steps mentioned above until the valid adjustment value is obtained 13 The will f...

Страница 20: ...ches the Occlusion Limit Pressure an alarm is triggered and stop infusion The time to alarm Occlusion is directly related with flow rate and occlusion limit pressure set Note Turning the potentiometer clockwise increase the Pressure limit to alarm Occlusion 3 9 7 Sound Volume adjustment The alarm sound volume can be adjusted by using a small flat screwdriver to rotate clockwise to increase or anti...

Страница 21: ...flash with alarm The infusion pump will stop infusion after 27 minutes and turn off automatically after 30 minutes When the AC power is reconnected the infusion pump will recharge the battery automatically 3 10 Trouble Shooting 3 10 1 Alarms and Alarm Codes Alarm code Possible causes Action No IV set installed or IV set not properly installed Check whether IV set is mounted to the specified positi...

Страница 22: ...e IV set or lower down the sensitivity of occlusion 1 Battery voltage is low When running on battery indicates that battery voltage is low Promptly connect to AC power or replace with same type battery 2 Battery failed When running on AC indicates that battery failed Check the connection of the battery or replace with same type battery Infusion finished 1 Change to new solution container and set t...

Страница 23: ...ber or the drop chamber is too sloping the liquid is not dripping but flowing through the sidewall of the chamber Pump Cassette not running 1 Defective IV set or installed improperly causing the pump blocked 2 Pump cassette can t run due to residues and dirty Abnormal pressure detection Refer to the solutions of eliminating the problem on alarm If failed again contact customer service IV type sett...

Страница 24: ...1 Daily check Checking items Checking Procedures Check whether the exterior of LP8 is damaged Check the exterior of LP8 Check whether the power cord is damaged Check the power cord Symptom Checking points Action Whether the battery is installed properly Reinstall the battery properly Press but can t turn on Whether the AC power is connected Connect to the AC power and turn the rear panel switch on...

Страница 25: ...et 2 After IV set is assembled close the door press message will disappear Performance of door sensor 1 Press over 0 5 seconds to turn on the pump 2 Open the pump door will light Closing the door the turns off the alarm Chapter IV Maintenance 4 1 Daily Maintenance EOG sterilization ultrasonic sterilization as usual as possible Avoid to clean with Organic solvent like diluent alcohol etc 4 1 1 Digi...

Страница 26: ...ttery Make sure to keep the battery fully charged before use 1 When LP8 has been stored for a long period the battery will require recharging 2 The battery will require recharging every 6 months 3 Normal battery life is 2 years 4 For safety replace with the specified type of battery Note Disposal of damaged batteries shall be handled in accordance with the local rules recycle if possible 4 2 1 Rec...

Страница 27: ...placement time may be shortened depending on conditions 4 4 Storage Store the DigiPump LP8 in the following environment after cleaning Ambient temperature 5 C 40 C 41 F 104 F Relative humidity 90 Atmospheric pressure 70KPa 106Kpa Do not store the DigiPump LP8 in the following places a Direct sun light or strong light b Direct wind from heating apparatus air conditioner ventilator or humidifier etc...

Страница 28: ... be read before the machine is used According to the safety classification of medical electrical and electronic equipment it is class II equipment According to classification of electrical and electronic equipment with regard to protection against electric shock it belongs to CF type equipments It is rainproof instrument ...

Страница 29: ... or power failure 2 Neglect abuse or misuse of equipment 3 Servicing of equipment by person s other than DIGICARE INC 4 Any unit opened or tampered with without prior authorization When returning a infusion pump for extended warranty repair you must first contact DIGICARE BIOMEDICAL TECHNOLOGY INC to receive a Returned Goods Authorization Number RGA that is to be clearly marked on top of the shipp...