Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 18 -
3.
The formula for adjustment is:
Note: “Volume Infused Display” is for “
”volume. In order to measure the “Actual Volume
Infused” you need to use a graduated glass or similar to measure the actual infusion volume by yourself.
The adjustment for IV set should be in the range of -30% to +30%.
3.9.6 Occlusion Valve Adjustment
Use one small flat screwdriver to rotate the potentiometer located on the back panel and adjust the
Occlusion Pressure level to alarm. The Pressure range is 40Kpa to 140Kpa divided in
8 levels.
Note:
When the pressure in the infusion set reaches the Occlusion Limit Pressure,
an alarm is triggered and stop infusion. The time to alarm Occlusion is directly related with flow rate
and occlusion limit pressure set.
Note: Turning the potentiometer clockwise increase the Pressure limit to alarm Occlusion
3.9.7 Sound Volume adjustment
The alarm sound volume can be adjusted by using a small flat screwdriver to
rotate (clockwise to increase or anticlockwise to decrease) the potentiometer
located on the back panel. (as shown in the picture)
Alternatively, no sound alarm can be obtained by pressing the MUTE
on the front panel. For security, the alarm sound will be activated again after 3 minutes.
Position
indication
1drop/Min
80drops/Min
Volume
Position
1
Time to Alarm
5 min. 37 s.
Time to Alarm
5.6 seconds
0.37mL
Position
8
Time to Alarm
18 min. 16 s.
Time to Alarm
10.5 seconds
0.7mL