Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 3 -
E. General Safety
E.1 – Indications
The Digipump
LP8 infusion pump is a device designed for continuous operation. It is intended for use
by person’s trained in professional health care. The operator must be thoroughly familiar with the
instructions in this manual before using the instrument.
E.2 – Warnings
• The Digipump
LP8 infusion pump was designed and tested accordingly to the EMC Standard
IEC601-1-2. However, the operator is responsible to verify if the infusion pump is been affected or
affecting others
electrical equipment. Equipments like electrocautery and image scanners can generate interference
and cause degradation of the Digipump
LP8 infusion pump performance. To avoid this
situation, it should be installed as far as possible of those equipment.
• The Digipump
LP8 infusion pump is NOT intended to be used during MRI (magnetic resonance
imaging).
• The Digipump
LP8 infusion pump is NOT suitable for use in the presence of a flammable anaesthetic
mixture with air or with oxygen or nitrous oxide. Do not operate this product in the presence of
flammable anesthetics. Explosion may result.
• Follow the instructions on this manual. Set the parameters appropriately and use only the IV set which
has been adjusted for the DigiPump LP8.
• Stop the device immediately in case of any abnormal functioning of the infusion pump.
• Keep the DigiPump LP8 far, at least 1 meter (3.0 feet) from other generating HF equipments such as
electrosurgical unit and others.
• Before using the infusion pump, make sure the AC voltage at your installation site is correct for the
AC rating for the device. If the rating is not correct, do not use the infusion pump and contact
Digicare Customer Service for help.
• Follow local regulatory recommendations for Battery disposal. Do not overheat or place the battery into
fire as it may cause leakage, burning or explosion.
• Do not tear or break battery sheath as it may cause explosion or chemical burning.
• Hold on the plug firmly when plugging in or pulling out the power cord. Do not touch the power plug
with wet hand.
• Do not share the same power outlet with other appliances.
• Do not disassemble the device or modify it.
• Perform daily check before use. Make sure all functions perform well if the infusion pump is not being
in use.
• Avoid vibration, collision or exposition to the direct sunlight or other strong light.
• Avoid the warm or wet air flow coming from heating or cooling apparatus such as air-conditioning,
ventilator, humidifier, etc.