3. IV set distortion
due to long time use
in the same tube
segment
1)
After the IV set has been used over 5-6 hours, we suggest reposition
the IV tubing down or up 20cm for the linear pump cassette act on a
new tube position.
2)
We suggest the replacement of the IV set when it has been used for
20 hours or more.
Table 3.4
Adjustment value of different surface tension liquid (only for reference)
Liquid
Adjustment value
Standard (low surface tension)
0%
Dextrose(10%)
+5%
Dextrose (50%)
+10%
TPN
+5~10%
TPN+ fat-soluble vitamin
+10~20%
1. Assume that you want to infuse dextrose (10%) solution. The corresponding adjustment value is
+5% (table
3.4).
The IV set adjustment value is +3% through “
3.9.4 Auto-adjustment for IV set
” or “
3.9.5
Manual-adjustment for IV set
”;
2. The total adjustment value = IV set adjustment value (3%) + High Surface Tension liquid adjustment
value (5%), so the value needed as adjustment factor shall be 8%.
3. Press“
”or“
” to place the cursor at
, the
flashes;
4. Press“
”or“
”, set 8% ,then press “
” .
Warning: Setting a wrong Adjustment Factor will reduce the specified accuracy.