Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 14 -
3.5 Setting the flow rate and volume to be infused
1. Flow rate setting:
press“
”or “
”to move the cursor to “
”or“
”; At this
time, “
”or“
”will light. Press “
”or“
”to set the digit. Move to next digit
pressing “
”or “
” and move the cursor on the digit you want to set. Continue with this
procedure until you finish the rate value setting, then press “
”.
2. Volume to be infused setting:
press “
”or “
”, point the cursor on “
”.
The“
”will light. Set the volume digits following the same procedure as above, then press
“
”.
3.
Rate and Volume Setting Range
IV set
Ml/H
Drops/Min
Volume Limit
Setting Range
10drops/ML
1.0-999.9
1.0-166.6
1-9999ml
15drops/ML
1.0-999.9
1.0-249.9
1-9999ml
20drops/ML
1.0-999.9
1.0-333.3
1-9999ml
60drops/ML
1.0-150.0
1.0-150.0
1-9999ml
4. Empty bottle mode:
When setting the Volume Limit, press the“
”. The volume setting
will indicate
and will go to the “Empty” bottle mode, then press”
”to confirm.
Pressing “
” start infusion and the pump will infuse all liquid in the bottle automatically.
5.
Volume Infused Display
:
This indicator display the total volume infused. To clear this
indicator press“
”or“
” and point the cursor on“
”.
The“
”will flash. Press“
”for 0.5 seconds. The device will sound a beep and reset
total volume indicator to “0”, then press
to confirm.
3.6 Starting Infusion
1. Make sure all settings are correct;
2. Make sure there is no bubble in the IV set , then perform venipuncture.;