Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 16 -
Notes:
Turning power off and back on, reset the total infused volume to “0” automatically. The
IV set, flow rate, adjustment value and volume to be infused (limit) will keep the last
programmed settings.
The maximum volume displayed is 9999ml. If the total volume exceeds this value the
display will stay showing 9999m.
4. After all settings for IV set, Flow rate, adjustment factor and volume are done, press “
” to
restart infusion.
3.9.3 Purge Function
This function can be used to fill in, cleaning IV set or remove the air from the IV set quickly. In
addition, it can also be used for high rate infusion in a short time such as first aid etc. Please follow the
steps below.
1.
In Stop mode, press“
” button for 1.0 second to perform the purge function.
2.
Just release the“
” to stop the purge function.
3.
Purge Rate: for 10, 15, 20 drops/ml, IV set purge rate is 600ml/h; for 60 drops/ml, IV set purge
rate is 150ml/h.
NOTE:
During purge function, air in line alarm will only trigger a visual alarm, and not the audible
alarm, and the infusion will not stop. The volume infused is not added to the Total Volume
Infusion indicator.
3.9.4 IV Set Auto-adjustment
In order to keep the maximum specified infusion accuracy, we recommend the use of a reliable standard
PVC IV set ( outer diameter is 3.4mm-4.5mm). Use 0.9% standard physiologic solution or pure water for
adjustment following the steps below:
1.
Hang the liquid container filled with 200ml (or more) physiologic solution or pure water.
2.
Close the IV set roller clamp and make sure that no leakage occurs.
3.
Insert the IV set drop chamber to the liquid container, place the IV set in the pump with the drop
detector attached correctly, then open the roller clamp.
4.
Connect to AC power and turn on the pump.
Содержание DigiPump LP8