Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 24 -
4.1.3 Finger pump cassette
Remove the finger pump cassette and wash it,
wipe it with dry cotton and mount back .
4.1.4 Tube clamp
1)
Loosen screw and remove the tube clamp.
2)
Wash the tube clamp in water.
Dry it well .Then assemble tube clamp and fasten the screw.
4.1.5 Drop detector
1) Wipe the drop detector with a cloth dampened with warm and a mild detergent solution.
2) Dry it with a dry cloth and mount it back.
4.2 Battery
Make sure to keep the battery fully charged before use.
1)
When LP8 has been stored for a long period, the battery will require recharging.
2)
The battery will require recharging every 6 months.
3)
Normal battery life is 2 years.
4)
For safety, replace with the specified type of battery.
Note
: Disposal of damaged batteries shall be handled in accordance with the local rules, recycle if
possible.
4.2.1 Recharging
When the Low battery alarm, connect with the AC power, make sure the AC power switch is on, and
the device will recharge automatically.
1
)
Press“
”0.5 seconds to turn on the power , when “
”light, “
” flashing, it starts to
charge;
2
)
Press“
”0.5 seconds to turn off the power, it continues to charge; In approx 16