Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 11 -
D
A
B
C
Note
: To make sure it can infuse normally, do not leave the chamber full of liquid or the pump will not
be able to detect the drops. At the same time, if the chamber is not filled with liquid at least 1/4, the
pump will easily alarm because of the bubble formation in the chamber.
3. Open the roller clamp to remove air from IV set tubing.
Then close the roller clamp.
Note:
Make sure the air filter is open and tubing is correctly
placed in the pump or it may cause occlusion and alarm.
4.
Open the door and pull the tube clamp (D) to the left to insert
the IV set.
5. Keep the IV set tube straight and insert it to the guides
A, B and C
6. Release the roller clamp;
Note:
Make sure it doesn’t drip in the drop until you start
the infusion.
3.2.3 Install the drop detector
Install the drop detector following the instructions below:
1. Place the drop detector at middle point above the liquid level and bellow the chamber drop point.
(As show in the picture)
Note:
Make sure to install the drop detector correctly according to the above instructions or it may cause
alarms and errors.
2. Keep drop chamber straight. If it is over sloping, it may cause alarms and errors.
Drop
Indicator light of
drop detected
: It
will light for each
drop detected
1/4-1/3
of liquid
Keep
the
drop
detector flat