Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 17 -
5.
Press the “AUTO ADJUST” key on the back panel, the display indicator
will
flash and a beep will sound indicating the adjustment mode has been entered.
6.
Press“
”or“
” to move the cursor to
and set the type of IV set.
7.
Press“
”or“
”to move the cursor to
and set the flow rate to 200ml/h.
8.
Press“
”or“
” to the cursor at
and set the volume limit to 200ml.
9.
Press“
”to start the auto-adjustment function, pay attention to the drop detector indicator light
whether it flashes at each drop detected.
10.
During the auto-adjustment procedure,
will flash to show it is in the process of
adjustment. When 200ml has been infused completely, the pump will stop automatically and alarm.
11.
The LCD will indicate “
” after successful auto-adjustment,
will flash and
the adjustment factor value will be displayed. You can press“
”to store the value. If the process
of adjustment does not run properly and stops (Open, Occlusion etc.), or if the adjustment value is
not in the range of -30% to +30%, the LCD will indicate “
”alarm to show the failure of the
adjustment process. You need to turn off the pump then turn on again and restart the adjustment
process.
12.
Check whether the solution container has been emptied. If not, the adjustment value will be useless.
Reset the machine and repeat the steps mentioned above until the valid adjustment value is obtained.
13.
The
will flash and the adjustment value (automatically calculated by system) is
displayed. Press
to confirm and store the result in the memory which will be used as default
setting on the next power on.
Note:
If the hospital uses different types of IV sets, you must get their corresponding adjustment
value according to the auto-adjustment function mentioned above. Write them down and set
the corresponding value when use, it will ensure the maximum infusion accuracy.
3.9.5 IV set Manual Adjustment
The user can calculate the error and manually enter the adjustment factor.
1.
To verify error in IV sets of different brand or type, use 0.9% standard physiologic solution or pure
water for infusion. Set the flow rate to 125ml/hrin (range: 100ml to 150ml/hr).
2.
After more than 50ml is infused, calculate the adjustment factor.