4 Transportation and installation
Operating Instructions inLab MC X5
4.4 Connecting the suction device
båÖäáëÜ
4.4
Connecting the suction device
This section describes the connection of the Sirona suction. If you use a
suction device from an external supplier, note the following information.
1. If you use the Sirona suction device, connect the hose supplied with
the suction device to the connection (A) on the back of the unit.
2. Connect the suction device and the unit using the communication
cable supplied with the suction device, so as to enable the automatic
mode. To do this, connect the communication cable to the device
over the interface "Communication interface for suction (E)".
3. If you have a central or another suction device, connect this to the
connection (A).
A
E
IMPORTANT
Follow the requirements
When using a suction device that is not approved by Sirona, follow the
requirements in section "Suction requirements [ → 28]“.
Ensure the correct configuration of the suction device in the machine
configuration menu.
Note that the pipe to the suction device must have an inner diameter of
38 mm.