4 Transportation and installation
Dentsply Sirona
4.5 Connecting the coolant tank
Operating Instructions inLab MC X5
4.5
Connecting the coolant tank
In order to connect the coolant system, proceed as follows:
1. Mount the filter cartridge of the tank on the filter cartridge of the tank.
2. Place the filter cartridge into the tank and lock the tank opening.
3. Connect the coolant tank to the coolant drainage connection (A)
using the coolant return hose supplied with the tank as well as the
accompanying angle unit.
IMPORTANT
Coolant-lubricant mix
Make sure that the configuration of the coolant-lubricant mix is correct
according to the material class to be processed:
- Tank contents material class sintering metal: 10 L water
- Tank contents material class PMMA/composite/grinding materials:
250ml De 10L water
- Tank contents material class metal: 1bottle of DentaLub (900 ml) +
10L water
- Tank contents cleaning tank: 10 L water
A