7 Maintenance and cleaning
Operating Instructions inLab MC X5
7.8 Procedure for processing sintering metal
båÖäáëÜ
7.8
Procedure for processing sintering metal
7.8.1
Important information
NOTICE
Observe the safety information from the material manufacturer
Observe the safety instructions regarding occupational safety and
disposal referred to in the material manufacturer's operating
instructions. Dirty filters, tanks and metal residue must be disposed of
appropriately in accordance with these specifications.
IMPORTANT
Use a waterproof receptacle with a nominal volume of 10-20l as a
collecting vessel for changing the water (e.g. a commercially available
10l or 20l plastic bucket). When selecting the collecting vessel, bear in
mind that it will also serve as a transporting container for disposal and is
not reusable.
IMPORTANT
Wearing protective waterproof gloves is recommended.
IMPORTANT
Sirona recommends, however, that you use an additional standard tank
which is only used for processing sintering metal materials. This is
available individually as “Coolant tank complete (REF: 63 85 137)” or in
combination with a new tool magazine and the accompanying sintering
metal grinder “Starter Kit Sintering Metal inLab MC X5 (REF 65 61 986)”.
IMPORTANT
Ask your disposal company whether the filter waste needs to be
correctly sorted for disposal.