![DENTSPLY inLab MC X5 Скачать руководство пользователя страница 101](http://html1.mh-extra.com/html/dentsply/inlab-mc-x5/inlab-mc-x5_operating-instructions-manual_2483578101.webp)
64 78 759 D3608
D3608
.
201.01.07
.
02 10.2017
101
Dentsply Sirona
7 Maintenance and cleaning
Operating Instructions inLab MC X5
7.9 Procedure when processing Medentika-PreFace® abutment semi-finished products made of
båÖäáëÜ
2. Open the tank lid clip (C) and pull the filter system out. Pay attention
to any remainder liquid.
3. Drain two-thirds of the coolant tank.
4. Shake the tank.
5. Drain the coolant tank.
6. Shake the contents of a bottle of DentaLub into the tank.
7. Fill the tank with water up to the mark (
max
, D) (approx. 10 liters).
8. Place the filter system in the tank.
9. Lock the tank with the tank lid clip.
10. Connect the tank inlet hose (A) with the large diameter as well as the
thin coolant supply hose (B) again to the tank system.
NOTICE
Foaming not permissible!
If any cleaning agents are used, this will create foam, which is not
permitted.
➢ Do not use any cleaning agents.