4 Transportation and installation
Dentsply Sirona
4.6 Connecting the filter regulator
Operating Instructions inLab MC X5
4.6
Connecting the filter regulator
1. Screw the filter regulator firmly at the corresponding place (A) on the
device using the screws provided.
Alternatively, the filter regulator can be positioned for improved
visibility of the print display even at an external position with
corresponding conditions.
2. Connect the left compressed air outlet (B) of the filter regulator with
the inlet on the device side. Use the accompanying compressed air
pipe for this (blue, 0.75m). Shorten the pipe by approx. 30 cm and
connect the left output point of the filter regulator with the machine
connection point.
3. Connect the right compressed air connection (C) of the filter regulator
with the stationary compressed air connection at the installation site.
To do this, use the compressed air pipe (3 m) as well as the provided
quick-connect couplings. If you have a special compressed air
connection system, contact your installer in charge.
NOTICE
Avoiding any damage to the machine
The unit is equipped with a high-frequency mandrel that has a
pneumatic collet chuck and an air purge mechanism. A filter regulator
that filters impurities from the compressed air and protects the machine
from any damage should be interposed for supply of the required
compressed air.