![DENTSPLY inLab MC X5 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/dentsply/inlab-mc-x5/inlab-mc-x5_operating-instructions-manual_2483578031.webp)
64 78 759 D3608
D3608
.
201.01.07
.
02 10.2017
31
Dentsply Sirona
4 Transportation and installation
Operating Instructions inLab MC X5
4.5 Connecting the coolant tank
båÖäáëÜ
4. Connect the coolant hose to the coolant inlet connection (C) for
supply of coolant to the machine.
5. Connect the coolant inlet hose as well as the coolant return hose to
the connections provided in the tank system.
6. Install the tank underneath the machine and establish a direct hose
supply for the coolant return hose without a dip. Remove the angle
unit and shorten the hose if required for this. If required establish a
direct connection between the tank and the machine via a hole in the
work set-up area.
C
A
A
C
B
D
NOTICE
Risk of overflowing
Overflows may occur if the coolant return hose of the tank sags.
➢ Mount the tank, so as to ensure continuous inclination of the coolant
return hose, particularly in the hose piece on the table surface
immediately after the outlet from the back side of the machine.
➢ Ideally, position the tank under the machine. You may have to turn
the tank or shift it to do this.