4 Transportation and installation
Dentsply Sirona
4.3 Installation conditions
Operating Instructions inLab MC X5
4.3
Installation conditions
4.3.1
Installation site requirements
Make sure that the device is erected on an even, horizontal surface of
approx. 640 x 600 mm (W x D). The maximum capacity should be at least
90 kg. The height of the machine is 795 mm.
Erect the machine such that the main switch can be easily accessed.
Make sure that the ventilation slots on both sides of the unit remain
unobstructed.
The distance between the back of the unit and the wall must at least be 2
cm.
Note the weight of 87 kg!
The unit must not be installed at sites with a high level of humidity or dust!
4.3.2
Suction requirements
We recommend Sirona suction (optional); REF 6385129 (230V),
6485929 (120V).
If a suction device other than that recommended by Sirona is to be used,
the following requirements must be met:
● Suction power: 120 to 160 m
3
/h
● Dust class: M
● HEPA-microfilter, filter class: H12
4.3.3
Filter regulator/ compressed air supply requirements
Compressed air supply is required for operation of the device. This must
meet the following minimum requirements:
● Flow rate: 50 standard liters/min
● Pressure: min. 7.0bar / 102psi
● Compressed air quality according to ISO 8573-1:2010, class 1.4.2
– Impurities: Solid particle size < 1 µm
– Water content: Pressure condensation point < +3 °C
– Total oil content: < 0.1mg/m
3
The air pressure is monitored by the machine during operation. The
milling program stops upon falling below the given values. The milling
program can be continued once the values are correct and the error
message was acknowledged.