25
24
Modify BIT 3 of SPEC 12 to "1"
Click [REZERO] to enter zero calibration and wait
Put the corresponding weights, and click the [REZERO] button to
enter the full-scale calibration
Remove the weight to complete the calibration and return to the
weighing interface
Note:
1. The DSW-100 instrument system can be fully calibrated with a standard
weight greater than 10% of the full weight
2. The calibration switch (SPAN SW) must be on
4.2
Full scale rate correction
Click [REZERO] button to save
Click the switch below the scale body to open the calibration
Put a heavy object on the weighing platform, such as 10.005,
the weight has deviation
REZERO
Turn on the calibration switch again. Long press [REZERO],
click [ ← ] [ ↑ ] [ ↑ ] to enter the calibration, display the
previous calibration Scaling parameters, such as 37.500
+
REZERO
UNIT
←
TARE
↑
TARE
↑
4. CALIBRATION OPERATION
4.1 Full scale calibration
4. CALIBRATION OPERATION
4.2 Full scale rate correction
Click [ ← ] Shift, [ ↑ ] Increment, can be modified to the
weight to be calibrated, such as 5kg