【BIT 3】Do not use
0:Allow
0:Allow
【BIT 1】Manual TARE CLEAR
【BIT 0】When the calibration permission switch is
closed, the weight inner code display is allowed
1:Forbid
1:Forbid
*The default is 1001, there is no negative weight prompt display, showing
the weight value; manual tare weight clearing is allowed; weight inner code
display is prohibited when the calibration switch is allowed to be closed
SPEC NO.09
SPEC NO.10
【BIT 3】 NO Tare Accumulation
【BIT 1】Reset after peeling
0:Forbid
0: Weight calibration mode
0: Allow
0: Forbid
0: Free weight full value setting
1: Allow
1: Full correction mode
1: Forbid
1: Allow
1: Fixed weight full value setting
【BIT 2】 No Tare Subtraction
【BIT 0】 System mode select
*The default is 1010, both are dual range display; dual range is automatically
converted to single range; fixed range is 3000
*The default is 1100, which allows tare weight increase operation; allows tare
weight decrease operation; allows zero reset operation after tare removal;
range setting mode selection is set by free weight full value
*The default is 0011, the weight display decimal places are three decimal
places (000.000)
*The default is 0001, then the full-scale calibration method is the heavy
calibration mode; the minimum weight division value is 2
【BIT 2/BIT 1/BIT 0】
Decimal point position on of scale
【BIT 3】CALIBRATION MODE
【BIT 3】Dual rang weight check by
0:Only the net weight is dual range
0:Single range
1:Both are dual range display
1:Dual range
【BIT 2】Dual range automatic conversion (fixed range)Multi interval (only for fix mode1)
SPEC NO.11
SPEC NO.12
000: No decimal
point
100: Four decimal
places(00.0000)
001: One decimal
(00000.0)
101: Five decimal
places(0.00000)
010:Two decimal
places(0000.00)
011: Three decimal
places(000.000 )
000: 1
100: 20
001: 2
101: 50
010: 5
110: 100
011: 10
111: Not used
【BIT 2/BIT 1/BIT 0】Weight minimum division value
3. EDIT SETTINGS
3.5 System function definition
3.5 System function definition
3. EDIT SETTINGS
SPEC NO.06
【BIT 3/BIT 2】Negative weight display
00: When the net
weight is less
than negative 2D,
"UF." is displayed
01: When gross
weight is less
than negative 2D,
"UF." is displayed
10: There is no
l o a d w e i g h t
prompt display,
showing the load
value
1 1 : I n v a l i d
setting
【BIT 1/BIT 0】Fix full capacity select mode(Only
when SPEC 10, bit0=”1”)
00: 6000
01: 15000
10: 3000
11: Invalid setting
0:Allow
0:Allow
0:Allow
1:Forbid
1:Forbid
1:Forbid
【BIT 3】After the weight unit is converted,
the weight unit symbol flashes to allow setting
【BIT 2】Do not use
【BIT 1】Allow counting operation
【BIT 0】Exceeding the zero point area at
startup allows the use of the original zero position
*The default is 0001, the weight unit symbol is allowed to flash after the
weight unit is converted; counting operation is allowed; the original zero is
forbidden to be used when the area exceeds the zero point when starting
SPEC NO.07
00: g
0:Allow
0:Allow
01: kg
10: lb
1:Forbid
1:Forbid
11: OUNCE
【BIT 0】Not Stable Hold
SPEC NO.08
*The default is 1101, then the base (when calibration) weight unit is OUNCE;
when the weight is not zero, zero setting is allowed; when the weight is
unstable, locking is prohibited
【BIT 3/BIT 2】Benchmark (when calibrated) weight unit (g/kg/lb)
【BIT 1】WHEN WEIGHT IS NOT ZERO, NO PRESE TARE OP.