19
18
4. RS232C interface data format, stop bit is 1 bit, data bit 7 or 8 bits are
optional. (SPEC 01,02 = SELECT .)
5. After the USB port is connected, the machine will automatically detect the
data input by the USB port. If there is data input, the data communication of
the USB port will be turned on
*The default BIT 3 is 1, use the original set weight unit
No default BIT 2/BIT 1/BIT 0, POS/ECR communication protocol selection
(can be switched by pressing [Tare key] when turning on)
3.4
System function entry
Use the tool to click the (SPAN SW) button at the bottom of the
screen to make the calibration switch on.
Press and hold the [REZERO] key for a while, and click [ ← ] [ ↑ ] [ ← ]
to enter the spec setting.
+
REZERO
UNIT
←
UNIT
←
TARE
↑
UNIT
←
TARE
↑
After clicking the [ ↑ ] key to enter the setting, the [ ← ] key becomes
a shift key and will increase "0" from right to left; the [ ↑ ] key
becomes an incremental key and increases "1" from right to left.
After the operation, click [REZERO] to save and enter the next
item; click [ON/OFF] to exit without saving.
REZERO
ON
OFF
3.5
System function definition
SPEC NO.04
SPEC NO.05
00: ± 2% FS
00: ± 2% FS
0:<50% FS
0:Allow
【BIT 3/BIT 2】Reset range selection
【BIT 3】Auto clear conduction
【BIT 0】TARE RANGE
【BIT 0】AUTO CLEAR TARE
01: ± 4% FS
01: ± 4% FS
0:Forbid
0: When the net weight is greater
than 5d
10: ± 10% FS
10: ± 10% FS
1:Allow
1: When the net weight is greater than 5d
and the gross weight is greater than 21d
1: <100% FS
1:Forbid
11: Unlimited
11: Unlimited
【BIT 1】NO PRESET TARE OP.
【BIT 2/BIT 1】SYSTEM ON START ZERO RANGE
*The default is 0100, the reset range is ±4%FS; preset tare weight is
prohibited; tare range <50% FS
*The default is 1101, then the automatic clearing conditions are when the net
weight is greater than 5d and the gross weight is greater than 21d; the initial
zero point is allowed to be within ± 10% FS; automatic tare weight removal
is prohibited
3. EDIT SETTINGS
3. EDIT SETTINGS
3.4
System function entry
3.5 System function definition