5 Troubleshooting
5-16
05/04 1006-0453-000
Low Paw
high
(5)
In-range Paw data
available
Peak airway pressure <
|P
min
| + 4 for 20
consecutive seconds.
P
min
is the baseline
pressure during a
breath.
Check the breathing circuit for leaks.
Check flow sensors hoses for leaks.
Perform flow sensor calibration.
Check SIB to CPU cabling.
Replace the SIB board.
Replace the CPU board.
Alarm condition becomes false or transition from
mechanical ventilation on to off.
Low V
E
high
(5)
In-range flow data
available and alarm
limit not = off
V
E
< low limit
Check ventilator settings and volume output.
Check the breathing circuit.
Perform flow sensor calibration.
Replace the flow sensors.
Check SIB to CPU cabling.
Replace the SIB board.
Replace the CPU board.
Alarm condition becomes false or volume alarms
state transition to standby (or off).
Low V
TE
medium
In-range flow data
available and alarm
limit not = off
V
TE
< low limit
Check ventilator settings and volume output.
Check the breathing circuit and flow sensor
connections.
Perform flow sensor calibration.
Replace the flow sensors.
Check SIB to CPU cabling.
Replace the SIB board.
Replace the CPU board.
Alarm condition becomes false or volume alarms
state transition to standby (or off).
Manifold
Pressure
Sensor
Failure
Minimum
Monitoring
(8) (9)
Cal failure at bootup or
DPL switch engaged and
P
manifold
< 80
cm H
2
O
Check the breathing circuit.
Perform flow sensor calibration. If calibration
fails, replace the SIB.
Check DPL switch for proper operation using
the service mode Test press. limit switch
check. If DPL switch test fails, replace the SIB.
High
Pressure
Limit
Reached
Minimum
Monitoring
(8) (9)
DPL engaged and Paw <
90
cm H
2
O
and
Pmanifold < 80
cm H
2
O
Check the breathing circuit.
Perform flow sensor calibration. If calibration
fails, replace the SIB.
Check the DPL switch for proper operation
using the service mode Test press. limit switch
check. If DPL switch test fails, replace the SIB.
Memory
(EEPROM)
Fail
User Alarm
The system cannot
access some stored
values.
Default settings are
used. Ventilation is still
possible but service is
necessary.
Replace the CPU board.
Memory
(flash)
Failure
(7)
Minimum
Shutdown
(6)
CRC failure
Replace the CPU board.
Message
Alarm type/
Priority
Enabling Criteria
Alarm Condition
Service Repair
Removal Criteria
Содержание Aestiva 7900 SmartVent
Страница 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Страница 44: ...Notes 2 28 05 04 1006 0453 000 ...
Страница 81: ...4a Tests and Calibration 4 X Software 1006 0453 000 05 04 4a 35 Figure 4 38 Service Calibration full menu ...
Страница 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Страница 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Страница 150: ...Notes 5 34 05 04 1006 0453 000 ...
Страница 158: ...Notes 6 8 05 04 1006 0453 000 ...
Страница 182: ...Notes 7 24 05 04 1006 0453 000 ...
Страница 189: ...8 Illustrated Parts 1006 0453 000 05 04 8 7 U14 U23 4 12 13 14 15 16 15 18 21 20 23 24 1 1a 22 17 19 ...
Страница 202: ...Notes 8 20 05 04 1006 0453 000 ...
Страница 203: ......