4a Tests and Calibration 4.X Software
1006-0453-000 05/04
4a-19
4a.3.10 Test 5V Fail
Alarm
A 5-Volt supply (VDD) is used to power the digital circuits throughout the
ventilator.
The 5-Volt supply (VDD) is derived in the power section of the control board. It
is used to power the digital circuits throughout the ventilator. If the 5-Volt
supply fails, the ventilator will sound a continuous alarm tone when the system
switch is turned on.
To test the 5V Fail Alarm, follow the directions on the screen.
If the alarm tone does not sound, replace the CPU board (see section 7).
Figure 4-20 • Test 5V Fail Alarm menu