
7 Repair Procedures
7-18
05/04 1006-0453-000
7.9 Gas inlet valve
1. Tube guide plate
2. GIV solenoid switch
3. Captive mounting screws (2)
4. GIV
5. Inlet filter housing
6. Manifold plate
Figure 7-16 • Gas Inlet Valve (GIV)
Follow the instructions in Section 7.6 to remove the Vent Engine.
To remove the solenoid switch:
1. Use a 1.5 mm hex wrench to remove the mounting screws from the GIV
solenoid switch on the inlet valve.
To replace the gas inlet valve:
1. Remove the tube guide plate from the top of the GIV.
2. Remove the inlet filter underneath the manifold plate.
3. Use a 3 mm hex wrench to loosen the two captive inlet valve mounting
screws.
4. Remove the GIV/inlet filter housing assembly.
5. To separate the GIV from the inlet filter housing, remove the two mounting
screws from within the filter housing.
6. Examine the valve seat and O-rings in the manifold for damage. Replace as
necessary.
7. Reassemble in reverse order.
8. Perform the Checkout Procedure found in Section 3 of the Aestiva
Anesthesia Machine Service Manual.
1
2
3
4
5
6
Содержание Aestiva 7900 SmartVent
Страница 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Страница 44: ...Notes 2 28 05 04 1006 0453 000 ...
Страница 81: ...4a Tests and Calibration 4 X Software 1006 0453 000 05 04 4a 35 Figure 4 38 Service Calibration full menu ...
Страница 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Страница 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Страница 150: ...Notes 5 34 05 04 1006 0453 000 ...
Страница 158: ...Notes 6 8 05 04 1006 0453 000 ...
Страница 182: ...Notes 7 24 05 04 1006 0453 000 ...
Страница 189: ...8 Illustrated Parts 1006 0453 000 05 04 8 7 U14 U23 4 12 13 14 15 16 15 18 21 20 23 24 1 1a 22 17 19 ...
Страница 202: ...Notes 8 20 05 04 1006 0453 000 ...
Страница 203: ......