4b Tests and Calibration 1.X/3.X Software
4b-22
05/04 1006-0453-000
4b.8 Adjust Drive Gas Regulator
The drive gas regulator must be calibrated under flow conditions.
The
Adjust Drive Gas Regulator
procedure establishes the required flow rate
through the drive gas regulator for proper calibration.
Remarks
The drive gas regulator should provide a constant gas input pressure of
172 kPa (25 psi).
You can verify this pressure by attaching a pressure test device to the
regulator pressure port (shown below) and performing the above procedure.
If required, adjust the regulator to 172 ±1.72 kPa (25 ±0.25 psi).
Regulator
pressure
port