4b Tests and Calibration 1.X/3.X Software
4b-16
05/04 1006-0453-000
Notes:
1. These columns show the acceptable range where possible.
2. The A/D count and displayed value shown for Airway Pressure, Manifold Pressure, Inspiratory Flow, and Expiratory Flow are the zero
pressure values. These readings should be taken with the flow sensors disconnected by removing the circuit module.
3. The count range and displayed value shown for GIV Current Sense is with the GIV turned on. If the GIV is turned off, the off count range
is 0 to 259 Counts.
Table 4b-1 A/D Channels
A/D Channel
Counts
1
Actual Range
1
Displayed Range
Special Instructions
Airway Pressure
2
550-1050
Near Zero
cm H
2
O
-20 to 120
cm H
2
O
Zero Offset Reading (nominal 800 Counts)
Manifold Pressure
2
550-1050
Near Zero
cm H
2
O
-20 to 120
cm H
2
O
Zero Offset Reading (nominal 800 Counts)
Inspiratory Flow
2
1800-2300
Near Zero L/min
-120 to 120 L/min
Zero Offset Reading (nominal 2050
Counts)
Expiratory Flow
2
1800-2300
Near Zero L/min
-120 to 120 L/min
Zero Offset Reading (nominal 2050
Counts)
Flow Current Sense
0-4095
0 to 1.024 Amp
0 to 1.024 Amps
Flow DAC Feedback
0-4095
0 to 4.095 Volts
0 to 4.095 Volts
GIV Current Sense
3
2371-3884
143.7 to 235.4 mA
0 to 248.2 mA
Off state reading is 0 to 259 counts
O2 Concentration
0-4095
0 to 100%
0 to 100%
Count weight and limits are determined
during O2 calibration
1.225 Voltage
Reference
1214-1235
1.214 to 1.235 Volts
0 to 4.095 Volts
5.8V Voltage Test
1997-2565
4.86 to 6.24 Volts
0 to 9.96 Volts
+15V Analog Supply
1914-2262
13.77 to 16.27 Volts
0 to 29.46 Volts
-15V Analog Supply
858-953
-12.62 to -17.46 Volts -61 to 1.43 Volts
+12V SIB Supply
1779-2322
10.44 to 13.62 Volts
0 to 24 Volts
VEL Supply
1645-2675
9.56 to 15.69 Volts
0 to 24 Volts
VSW Supply
1269-2368
9.13 to 17.04 Volts
0 to 29.46 Volts
+12V Light Supply
1879-2217
11.02 to 13 Volts
0 to 24 Volts
In Rev 3.4 software,
Range = 10.00 to 12.73 Volts
Battery Voltage
1740-3976
7 to 16 Volts
0 to 16.48 Volts
<7V = Low battery voltage failure.
<11.65V = Low battery voltage warning.
>16 Volts (10 Sec) = High battery voltage
failure.
Battery Current
700-4000
-600 mA to 6 Amps
-2 to 6.19 Amps
-150 to -600 mA = Battery charging.
-601 mA to -2 Amps = Charge current out of
range.
>300 mA = Operating on battery.
>6 Amps = Battery discharge current out of
range.
Содержание Aestiva 7900 SmartVent
Страница 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Страница 44: ...Notes 2 28 05 04 1006 0453 000 ...
Страница 81: ...4a Tests and Calibration 4 X Software 1006 0453 000 05 04 4a 35 Figure 4 38 Service Calibration full menu ...
Страница 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Страница 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Страница 150: ...Notes 5 34 05 04 1006 0453 000 ...
Страница 158: ...Notes 6 8 05 04 1006 0453 000 ...
Страница 182: ...Notes 7 24 05 04 1006 0453 000 ...
Страница 189: ...8 Illustrated Parts 1006 0453 000 05 04 8 7 U14 U23 4 12 13 14 15 16 15 18 21 20 23 24 1 1a 22 17 19 ...
Страница 202: ...Notes 8 20 05 04 1006 0453 000 ...
Страница 203: ......