2 Theory of Operation
1006-0453-000 05/04
2-5
In current Aestiva machines, the regulated power supplies and the CPU are
included on a single board (Integrated CPU Board), as shown in Figure 2-3.
A universal power supply is used to convert AC to DC that feeds into the power
supply circuits of the integrated CPU board. The universal power supply
eliminates the need for the toroidal transformer.
Figure 2-3 • Aestiva 7900 Ventilator functional block diagram (Integrated CPU)
AB.29.082
-Fuses
-Voltage Selector
-Filter
Connector Board
Connector Board
100-120 VAC
220-240 VAC
50/60 Hz
AC Inlet Module
Universal
Power Supply
Battery
Power Circuits
Digital Circuits
Front Panel
Assembly
Pneumatic
Engine
Sensor
Interface Board
Proprietary RS232C
Primary RS232C On/StandBy
Vent On/StandBy Switch
O2 Supply Switch and O2 Flush Switch
Patient Interface Harness
ACGO
Tec6 Outlet
Task Light
BCID
Oxygen Sensor
Integrated
CPU Board
Содержание Aestiva 7900 SmartVent
Страница 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Страница 44: ...Notes 2 28 05 04 1006 0453 000 ...
Страница 81: ...4a Tests and Calibration 4 X Software 1006 0453 000 05 04 4a 35 Figure 4 38 Service Calibration full menu ...
Страница 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Страница 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Страница 150: ...Notes 5 34 05 04 1006 0453 000 ...
Страница 158: ...Notes 6 8 05 04 1006 0453 000 ...
Страница 182: ...Notes 7 24 05 04 1006 0453 000 ...
Страница 189: ...8 Illustrated Parts 1006 0453 000 05 04 8 7 U14 U23 4 12 13 14 15 16 15 18 21 20 23 24 1 1a 22 17 19 ...
Страница 202: ...Notes 8 20 05 04 1006 0453 000 ...
Страница 203: ......