4b Tests and Calibration 1.X/3.X Software
4b-32
05/04 1006-0453-000
4b.12.5 V
E
Alarm Limits
The setting on the
V
E
Alarm Limits
menu determines how the
V
E
alarm limits
are set:
• If
V
E
Alarm Limits
is set to
Automatic
, the ventilator software calculates the
high and low
V
E
alarm limits and sets them to ±20% of the set
V
E
. The user is
still able to change the limits through the Alarm Settings selection in the
Main Menu.
• If
V
E
Alarm Limits
is set to
User Adjustable
, the user sets the limits on the
Alarm Settings
menu.
4b.12.6 User Select
Defaults
This menu (feature) is only available in 3.X software.
The
User Select Defaults
menu determines the control settings used by the
system at power up:
• If
last set
is selected, the system saves its settings when you turn it off and
powers up with the same settings.
• If
facility defaults
is selected, the system powers up with default facility
settings and returns to default settings when the user selects “End Case”.
Use this page to set the defaults as shown.
Screen with “last set” selected:
When
facility defaults
is selected, the screen expands to two pages:
• Page 1 of 2 — control settings
• Page 2 of 2 — alarm limits.
Содержание Aestiva 7900 SmartVent
Страница 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Страница 44: ...Notes 2 28 05 04 1006 0453 000 ...
Страница 81: ...4a Tests and Calibration 4 X Software 1006 0453 000 05 04 4a 35 Figure 4 38 Service Calibration full menu ...
Страница 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Страница 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Страница 150: ...Notes 5 34 05 04 1006 0453 000 ...
Страница 158: ...Notes 6 8 05 04 1006 0453 000 ...
Страница 182: ...Notes 7 24 05 04 1006 0453 000 ...
Страница 189: ...8 Illustrated Parts 1006 0453 000 05 04 8 7 U14 U23 4 12 13 14 15 16 15 18 21 20 23 24 1 1a 22 17 19 ...
Страница 202: ...Notes 8 20 05 04 1006 0453 000 ...
Страница 203: ......