6 Maintenance
1006-0453-000 05/04
6-7
6. Occlude the top of the standpipe tube with your finger, making sure the
tube remains 1/2 inch (15mm) above the blue seal.
7. Check the manifold pressure reading on the service screen. The MOPV
valve is functioning correctly if the steady state pressure reading is
<120cm H
2
O.
8. Set the system switch to Standby.
9. Remove the test plug from the exhalation valve port.
10. Remove the standpipe from the port and reinstall in the exhalation valve.
11. Reassemble the breathing system.
12. Perform the Preoperative Checkout Procedure in the Aestiva Operation
Manual.