4a Tests and Calibration 4.X Software
1006-0453-000 05/04
4a-21
4a.3.12 Breathing
System Leak Test
You can estimate how much of a leak there is in the ventilator portion of the
breathing system by closing the patient circuit, inflating the bellows, and
observing how quickly they fall on their own weight (part of the machine
checkout procedure).
The Breathing System Leak Test allows you to more precisely test the
ventilator portion of the breathing circuit for leaks.
By using the patient circuit to establish a closed loop, you can measure the
leak rate.
• The leak rate is the fresh gas flow needed to maintain 30 cmH2O.
• The system should have a leak rate <200 mL/min.
Figure 4-22 • Breathing System Leak Test menu
Содержание Aestiva 7900 SmartVent
Страница 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Страница 44: ...Notes 2 28 05 04 1006 0453 000 ...
Страница 81: ...4a Tests and Calibration 4 X Software 1006 0453 000 05 04 4a 35 Figure 4 38 Service Calibration full menu ...
Страница 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Страница 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Страница 150: ...Notes 5 34 05 04 1006 0453 000 ...
Страница 158: ...Notes 6 8 05 04 1006 0453 000 ...
Страница 182: ...Notes 7 24 05 04 1006 0453 000 ...
Страница 189: ...8 Illustrated Parts 1006 0453 000 05 04 8 7 U14 U23 4 12 13 14 15 16 15 18 21 20 23 24 1 1a 22 17 19 ...
Страница 202: ...Notes 8 20 05 04 1006 0453 000 ...
Страница 203: ......