7100 Anesthesia Ventilator
5-4
02/03 1006-0836-000
5.3 Alarm and Error messages
The Service Mode (refer to section 4) includes a log of the 20 most recent
Alarm messages (not shaded) and Error messages (shaded) experienced by
the ventilator system.
If a User Alarm persists after the recommended action has been performed,
the message indicates the probable component and related circuit that needs
repair. Use the Service Mode tests to isolate the fault. The items in the Service
Repair column indicate the path from the named component to the control
board.
Alarm Message
Error Message
Message
Alarm/Error
Cause
User Action/Concerns
Service Repair
12 Hour Test
w
User Alarm
System in use for more than
12 hours without a
powerup self test.
At end of case, move the system
switch from On to Standby to
On.
+12V Analog
Out-of- Range
Error
Minimum
Shutdown
Ventilator malfunction.
Ventilate manually.
Monitoring is not reliable.
Replace:
• Control Board.
-12V Analog
Out-of- Range
Error
Minimum
Shutdown
Ventilator malfunction.
Ventilate manually.
Monitoring is not reliable.
Replace:
• Control Board.
12V -10VA Test
Out-of-Range
Error
Minimum
Shutdown
Ventilator malfunction.
Ventilate manually.
Monitoring is not reliable.
Replace:
• Control Board.
Absorber Panel Open User Alarm
The top panel on Aestiva is
not completely closed.
ABS on Aespire in not
completely engaged.
Close the panel.
Reinstall ABS.
If persists, check:
• Panel switch
• Harness to MIA/
VMB
• MIA/VMB
• Cable to CB/J12
A/D Converter
Failure
Error
Minimum
Shutdown
Ventilator malfunction.
Ventilate manually.
Monitoring is not reliable.
Replace:
• Control Board.
Apnea Alarm
Standby
User Alarm
Normal condition after End
Case, powerup, or ACGO
change from On to Off.
Monitoring resumes after first
breath (mechanical) or 2
breaths within 30 sec (non-
mechanical).
Apnea Alarm Off
User Alarm
The cardiac bypass option is
selected (Alarm Settings
menu).
Apnea alarms are turned off
when this option is selected.