4 Tests and Calibration
1006-0836-000 02/03
4-31
4.10.7
Test EEPROM
When you start the
EEPROM Test
, the procedure performs the test once and
notes whether the test passed or failed.
If the EEPROM test fails, replace the control board.
EEPROM Test
The software tests all of the EEPROM memory via a bit pattern test. It writes a
certain pattern to a block of memory and then reads that block of memory. If
the bit pattern that was written is not the same as the bit pattern read back, the
test fails.