5 Troubleshooting
1006-0836-000 02/03
5-15
5.4 Troubleshooting Flowcharts
5.4.1 Ventilator
assessment process
* Refer to Section 7.2,
“Breathing System Leak Tests,”
in the respective anesthesia machine Technical Reference Manual.
Start
Turn on System
Review Active Alarms,
Error Log and Alarm Log
No
Proceed
to “No Display”
Troubleshooting
Section 5.4.2
No
Use Alarm
Troubleshooting
Section 5.3
Proceed to
“Breathing System Leak”
Troubleshooting
Yes
No
Yes
Is display
on?
Are
there
alarms?
Diagnosis
Achieved?
Done
Alarms of Type:
Patient Circuit Leak
System Leak
Bellows Falls or
Leak detected
ALARMS
SYMPTOMS
Yes
Yes
No
Proceed to
“Inaccurate Volume
Ventilation”
Troubleshooting
Section 5.4.3
Yes
Alarms of Type:
Vt Compensation Off
Delivered Volume Mismatch
Check Flow Sensors
Insp Reverse Flow
Exp Reverse Flow
System Leak
Inaccurate Ventilation
Yes
No
Proceed to
“No Ventilation”
Troubleshooting
Section 5.4.4
Yes
Alarms of Type:
Low Paw
Volume Apnea
Low Drive Gas
Pressure System
Leak
Will not
ventilate
Yes
No
Proceed to
“High Intrinsic PEEP”
Troubleshooting
Section 5.4.5
Yes
Alarms of Type:
Sustained Paw
High Paw
Continuously high PEEP
during operation
Yes
No
No
No
Yes
*
AB.43.100