7100 Anesthesia Ventilator
7-20
02/03 1006-0836-000
7.5 Serial Adapter Board (SAB) and Power Cord/Harness
Note:
This section applies only to an
Aestiva machine
. For an
Aespire
machine
, refer to the Aespire Technical Reference Manual for details about the
Serial Isolation and Connector Board.
The SAB is located in the electrical enclosure area behind the AC Inlet module.
The two cables that connect to the MIA are behind the rear panel in the
pneumatics area.
w
WARNING
Disconnect the power cord from the wall before attempting to remove or
repair any circuit board to avoid shock hazard.
w
WARNING
The AC inlet module is very heavy. Use caution when removing it from the
anesthesia enclosure.
w
w
w
w
Caution
The circuit boards are electrostatic sensitive. Use an anti-static
workstation and wear a wrist grounding strap when handling a circuit
board.
To gain access to
the Serial Adapter Board
1. Remove the AC inlet module from the rear of the Aestiva Anesthesia
Machine.
a. Loosen the two captive M4 screws.
b. Pulling on the two captive screws, cautiously slide the AC inlet mod-
ule out only halfway from the enclosure. Use the two side handles
(cutouts in the sheet metal) to lift out the AC inlet module from the en-
closure.
w
WARNING
The AC inlet module is very heavy. Use caution
when removing it from the anesthesia enclosure.
.
2. Remove the electronics compartment cover by removing the four Phillips
head screws (two along top edge, one along each side).
3. Remove the rear panel
(see Section 4.1.1 of the Aestiva Anesthesia Machine Service Manual).
Captive screws
Cutout handles