5 Troubleshooting
1006-0836-000 02/03
5-9
Low Paw
User Alarm
Paw does not rise at least
4 cm above Pmin during
the last 20 sec.
Are circuit connections OK?
Look at the Paw gauge on the
absorber.
Look for circuit disconnection.
Low
◊
E
User Alarm
Exhaled minute volume
less than low limit alarm
setting. This alarm is
suspended for 9 breaths
after you change the
ventilator settings.
Check patient condition.
Check tubing connections.
Check alarm settings.
Low V
TE
User Alarm
Exhaled tidal volume
less than low limit alarm
setting. This alarm is
suspended for 9 breaths
after you change the
ventilator settings.
Check patient condition.
Check tubing connections.
Check alarm settings.
Memory (EEPROM)
Fail
User Alarm
The system cannot access
some stored values.
Default settings are used.
Ventilation is still possible but
service is necessary.
Replace:
• Control Board.
Memory (Flash)
Failure
Error
Minimum
Shutdown
Ventilator malfunction.
Ventilate manually.
Monitoring is not reliable.
Replace:
• Control Board.
Memory (RAM)
Failure
Error
Minimum
Shutdown
Ventilator malfunction.
Ventilate manually.
Monitoring is not reliable.
Replace:
• Control Board.
Memory (Redundant
Storage) Failure
Error
Minimum
Shutdown
Ventilator malfunction.
Ventilate manually.
Monitoring is still available.
Replace:
• Control Board.
Memory (Video)
Failure
Error
Minimum
Shutdown
Ventilator malfunction.
Ventilate manually.
Monitoring is not reliable.
Replace:
• Control Board.
Monitoring Only
User Alarm
A severe malfunction
prevents mechanical
ventilation.
Other alarms may also
occur.
Ventilate manually. Cycle system
power (On- Standby-On).
If the alarm clears, restart
mechanical ventilation.
Replace:
• Control Board.
No Circuit Module
User Alarm
The patient circuit module is
not installed
(Aestiva only).
Install a module.
Optical sensors look for tabs on
the back of the module.
Is the module assembled?
Are sensors dirty?
If persists, check:
• Circuit
Identification
Board
• Harness to MIA/
VMB
• MIA/VMB
• Cable to CB/J12
Message
Alarm/Error
Cause
User Action/Concerns
Service Repair