4 Tests and Calibration
1006-0836-000 02/03
4-25
Notes:
1. The
Counts
column shows typical digital counts from the computer with the expected (
range
) shown in parenthesis.
2. The
Actual
column shows current values in real time with inputs in their baseline state.
A/D Channel
Counts
1
(range)
Actual
2
Range
Comments
Inspiratory Flow
2050
(1800-2300)
0.000 L/min
-120 to 120 L/min
Zero Offset Reading (nominal 2050 Counts)
Increased flow = more counts/more positive
Expiratory Flow
2050
(1800-2300)
0.000 L/min
-120 to 120 L/min
Zero Offset Reading (nominal 2050 Counts)
Increased flow = less counts/more negative
Airway Pressure
800
(550-1050)
0.000 cmH2O
-20 to 120 cmH2O
Zero Offset Reading (nominal 800 Counts)
O
2
377
(0-4095)
21.000%
0 to 100%
Count weight and limits are determined during
O
2
calibration
Flow Valve Volt
0
(0-4095)
0.000 Volts
0 to 6.140 Volts
PEEP Valve Volt
0
(0-4095)
0.000 Volts
0 to 6.140 Volts
Flow DAC Feedback
0
(0-4095)
0.000 Volts
0 to 4.095 Volts
PEEP DAC Feedback 0
(0-4095)
0.000 Volts
0 to 4.095 Volts
1.225 Volt Ref
1.225
(1214-1235)
1.225 Volts
1.214 to 1.235 Volts
+12V 10VA Supply
2041
11.975 Volts
10.5 to 13.5 Volts
12V-10VA Test Out-of-Range
(Minimum Shutdown)
+12V Analog Supply 2050
12.027 Volts
10.5 to 13.5 Volts
+12V Analog Out-of-Range
(Minimum Shutdown)
-12V Analog Supply
396
-12.071 Volts
-10.2 to -13.86 Volts
-12V Analog Out-of-Range
(Minimum Shutdown)
Battery Voltage
3609
7.215 Volts
0 to 8.220 Volts
> 8 Volts (10 sec) = Battery Failure High
< 5.65 Volts = Low Battery Voltage
< 2 Volts = Battery Failure Low
Battery Current
997
-0.008 Amps
-2 to 6.190 Amps
< -750 mA = Battery Charger Fail
-330 to -700 mA (1 min) = Battery charging
w
>600 mA (10 sec)= On Battery-Power OK?
>6 Amps (10 sec) = Battery Current High