4 Tests and Calibration
1006-0836-000 02/03
4-27
4.10.3
Display Battery Status
The
Display Battery Status
menu displays the battery charge status.
Battery Status
(on line power)
“Battery Charged” — If none of the following conditions are in effect.
“Battery Charging
w
” — Battery Current = -330 to -700 mA
“Battery Failure Low” — Battery Voltage <2 Volts
“Battery Failure High” — Battery Voltage >8 Volts
“Battery Current High” — Battery Current >6 Amps
“Battery Charger Fail” — Battery Current < -750 mA
Battery Status
(on battery power)
“On Battery-Power OK?” — System is running on battery.
“Low Battery Voltage” — Battery Voltage <5.65 Volts
Remarks
While the battery is charging, the battery current is displayed as negative
current. If the battery has been on charge for a long time (8 hours minimum)
and you do not get a “Battery Charged” display:
•
The battery has failed and you should replace it.
When the battery is discharging, the battery current is displayed as positive
current. Disconnect the power cord and observe the discharge current.
•
A good, fully-charged battery should maintain a steady discharge
current (for at least 30 minutes under normal operation).
•
If the discharge current and battery voltage drops off quickly, the
battery has failed and you should replace it.
w
w
w
w
Warning
Depending on the battery condition and the ventilator settings,
continued operation of the mechanical ventilation may result in sudden
shut down.