4 Tests and Calibration
1006-0836-000 02/03
4-33
4.10.9
Test 5V Fail Alarm
The 5-Volt supply (VDD) is derived in the power section of the control board. It
is used to power the digital circuits throughout the ventilator. If the 5-Volt
supply fails, the ventilator will sound a continuous alarm tone when the system
switch is turned on.
To
Test the 5V Fail Alarm
, follow the directions on the screen.
Remarks
If the alarm tone does not sound, replace the control board.